Abstract
Quantitative scanning transmission electron microscopy (STEM) allows composition determination for nanomaterials at an atomic scale. To improve the accuracy of the results obtained, optimized imaging parameters should be chosen for annular dark field imaging. In a simulation study, we investigate the influence of imaging parameters on the accuracy of the composition determination with the example of ternary III-V semiconductors. It is shown that inner and outer detector angles and semi-convergence angle can be optimized, also in dependence on specimen thickness. Both, a minimum sampling of the image and a minimum electron dose are required. These findings are applied experimentally by using a fast pixelated detector to allow free choice of detector angles.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.