Abstract

Abnormal slow-wave activity has been associated with functional motility disorders. Gastric pacing has been investigated to correct slow-wave abnormalities, but clinical therapies are yet to be established. This study aimed to define optimal parameters to advance the application of gastric pacing. High-resolution gastric mapping was utilized to evaluate four pacing parameters in in-vivo pig studies: (i) orientation of the pacing electrodes (longitudinal vs circumferential), (ii) pacing energy (900 vs 10,000 ms mA2), (iii) the pacing location (corpus vs antrum), and (iv) pacing period (between 12 and 36 s). The probability of slow-wave initiation and entrainment with the pacing electrodes oriented longitudinally was significantly higher than with electrodes orientated circumferentially (86 vs 10%). High-energy pacing accelerated entrainment over the entire mapped field compared to low-energy pacing (3.1±1.5 vs 7.3±2.4 impulses, p < 0.001). Regardless of the location of the pacing site, the new site of slow-wave initiation was always located 4-12 mm away from the pacing site, between the greater curvature and negative pacing electrode. A pacing period between 14-30 s resulted in stable slow-wave initiation and entrainment. These data will now inform effective application of gastric pacing in future studies, including human translation.