Abstract

AbstractAn electrolyte‐insulator‐semiconductor capacitor (EISCAP) sensor with the unique differential design is proposed where a liquid‐free reference (LR) device is used as the reference signal source. After the output matching based on AC excitation adjustment, LR device can provide up to 80 % elimination efficiency of background signal for differential EISCAP sensor. Experiment results of pH sensing demonstrate that the differential EISCAP sensor has the better accuracy and time stability compared with conventional EISCAP sensor. The differential design provides a feasible optimization direction for EISCAP sensor in monitoring applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.