Abstract
Radioscintigraphic imaging during sentinel lymph node (SLN) mapping could potentially improve localization; however, parallel-hole collimators have certain limitations. In this study, we explored the use of coded aperture (CA) collimators. Equations were derived for the six major dependent variables of CA collimators (i.e., masks) as a function of the ten major independent variables, and an optimized mask was fabricated. After validation, dual-modality CA and near-infrared (NIR) fluorescence SLN mapping were performed in pigs. Mask optimization required the judicious balance of competing dependent variables, resulting in sensitivity of 0.35%, XY resolution of 2.0 mm, and Z resolution of 4.2 mm at an 11.5-cm field of view. The findings in pigs suggested that NIR fluorescence imaging and CA radioscintigraphy could be complementary, but present difficult technical challenges. This study lays the foundation for using CA collimation for SLN mapping, and also exposes several problems that require further investigation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.