Abstract

A chemiluminescent nitrogen detector was optimized for packed-column supercritical fluid chromatography. Methanol modifier concentrations of 15% or less, an ozone flow of 5.8 mL/min, and a decompressed CO(2) flow between 240 and 310 mL/min were found to exhibit a maximum sensitivity of 5 ng for sulfamethazine (1 ng of N). The addition of a membrane drier to the "optimized" system further decreased the minimum detectable quantity (MDQ) to 0.5 ng (0.1 ng of N). In addition, by using a microbore column (2 mm i.d.) instead of an analytical scale column, the postcolumn decompressed flow split could be eliminated, further reducing the MDQ to 0.125 ng of sulfamethazine (0.025 ng of N).