Abstract

A new method for uranium (U) determination in presence of plutonium (Pu) in the range of 0.7–2.4 mg per aliquot is optimized in a low volume potentiometric cell (4 mL) and waste volume is decreased by six times. U determination in presence Pu at higher dynamic range is optimized by a complexometric titration by employing 2,6 pyridine di-carboxylic acid and ethylene diamine tetra acetic acid as the masking agent. In this method Pu introduced a bias of 6 % irrespective of U–Pu concentration ratio with a precision of 0.4 %.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.