Abstract

A novel method for preparing thin films was investigated for quantifying gallium and iron in plutonium solutions using WDXRF. This technique was developed to eliminate the potential for radioactive liquid to leak into the spectrometer, decrease specimen preparation time, and minimize waste. Samples were cast in μL quantities onto Kapton, and a surfactant was added to disperse the solution uniformly across the Kapton. After drying the specimens, they were sealed in a cell for analysis. Results to date indicate the method can provide a relative precision of ∼0.5% for gallium and ∼2% for iron, which is more than sufficient for routine sample analyses.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.