Abstract
Preparing dry specimens from liquid samples for XRF analysis avoids introducing caustic or hazardous liquids into the instrument. Several modifications were made to a dried residue specimen preparation method for quantifying gallium in plutonium metal in order to improve the method accuracy and precision. Ion exchange chromatography was utilized to remove the plutonium prior to casting the dried residue specimens. This, coupled with several other changes, improved the method relative error from ∼5% to less than 1%. These results are sufficient for routine sample analysis and are almost comparable to results from the established process using liquid specimens. However, the analysis of radioactive liquid specimens is unnecessary for quantifying the plutonium gallium content using this dried residue approach
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.