Abstract
We have developed a rapid and sensitive GC–MS assay for methylmalonic acid determination in serum and plasma utilizing an anion exchange solid-phase extraction and trimethylsilyl derivatization. Each step of the procedure was optimized by the experimental design methods to assure the assay reliable performance. The limit of detection and limit of quantitation were 0.025 and 0.1 μmol/l. The total coefficient of variation for the method was 9.8, 4.4, and 4.6% at the concentration of 0.2, 3.1, and 6.2 μmol/l methylmalonic acid concentration, respectively. The assay are linear up to 9.0 μmol/l, and showed good correlation with a reference method. The method has proven to be reliable in routine production, producing clean chromatography, unique ion fragments, and consistent ion mass ratio.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.