Abstract

The LabSOCS software was used to optimise the geometric dimensions of cylindrical containers and improve the detection efficiency of a high-purity germanium detector for the measurement of environmental radioactivity samples. The optimum ratio of diameter and height for different sample volumes were determined. In addition, the degrees of influence of the sample volume and density when the amount of sample was fixed were compared. Results show that the effect of sample volume on detection efficiency was significantly greater than that of sample density for a given sample amount.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.