Abstract
A computer model that incorporates the monochromatic aberrations of the eye is used to determine the optimal pupil size for axial and lateral resolution as it applies to retinal imaging instruments such as the confocal scanning laser ophthalmoscope. The optimal pupil size for axial resolution, based on the aberrations of 15 subjects, is 4.30 mm +/- 1.19 mm standard deviation (sd), which is larger than that for lateral resolution [2.46 mm +/- 0.66 mm (sd)]. When small confocal pinholes are used, the maximum detected light is obtained with a pupil size of 4.90 mm +/- 1.04 mm sd. It is recommended to use larger pupil sizes in imaging applications where axial resolution is desired.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Optical Society of America. A, Optics, image science, and vision
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.