Abstract
To compare a handheld and a digital pupillometer in determining pupil size in a population of refractive surgery candidates (group 1) and after implantation of an Artisan phakic intraocular lens (PIOL) for correction of myopia (group 2). Pupil size was measured with the Colvard and Procyon pupillometers in 121 eyes in group 1 and 83 eyes in group 2. Pupil sizes measured with the Colvard device were compared with the scotopic, mesopic-low and mesopic-high measurements taken with the Procyon pupillometer in both groups. Analysis of comparison between pupil measurements was performed according to methods described by Bland and Altman. The mean Colvard scotopic pupil diameter, scotopic, mesopic-low and mesopic-high Procyon pupil diameters were 5.86 +/- 0.81 mm, 6.42 +/- 0.88 mm, 5.55 +/- 0.95 mm and 4.21 +/- 0.73 mm in group 1 and 5.32 +/- 0.67 mm, 6.14 +/- 0.81 mm, 5.33 +/- 0.78 mm and 4.02 +/- 0.55 mm in group 2, respectively. The Colvard diameter compared most favourably with the Procyon mesopic-low diameter (group 2; p = 0.78). Measurements of pupil diameter with the Colvard pupillometer correlated best with measurements taken by the Procyon pupillometer under standardized mesopic-low light conditions. We believe that digital binocular infrared pupillometry is advantageous for obtaining standardized measurements of pupil size.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.