Abstract

When combined with on-line separations (e.g., capillary liquid chromatography (LC)), Fourier transform ion cyclotron resonance mass spectrometry (FTICR-MS) provides a powerful tool for biological applications, and particularly proteomic studies. The sensitivity, dynamic range, and duty cycle provided by FTICR-MS have been shown to be increased by ion trapping and accumulation in a two-dimensional (2D) radio-frequency (rf)-only multipole positioned externally to an FTICR cell. However, it is important that ions be detected across the desired m/z range without a significant bias. In this work we found that pressure inside the accumulation rf-quadrupole plays an important role in obtaining "unbiased" ion accumulation. Pressure optimization was performed in both pulsed and continuous modes. It was found that unbiased accumulation in a 2D rf-only quadrupole could be achieved in the pressure range of 5 x 10(-4) to 5 x 10(-3) Torr. External ion accumulation performed at the optimal pressure resulted in an increase in both the spectrum acquisition rates and dynamic range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.