Abstract

AbstractThe effect of stress distribution due to the changes of the distal screw alignment in relation to the Gamma nail and the femoral shaft is thoroughly studied in this paper. Failure of the Gamma nail composite occurs through the cranial aperture of the distal screws and the insertion hole for the lag screw due to nonunion, delayed-union and continued weight-bearing. A three-dimensional finite element model was used to study the fractured femur, the Gamma nail, the lag screw and the distal locking screws. The first and the second distal screws were inserted into the Gamma nail in four different configurations. We found that the stress of the Gamma nail composite was substantially reduced with the two screws configured in the anterior to posterior direction. This alignment can bear greater loading in the more demanding fracture types. In the subtrochanteric fracture or the comminuted fractures at the proximal femur, the optimal alignment of the two distal screws was in the anterior to posterior direction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.