Abstract
Optimal beam quality for detection of pulmonary nodules in digital chest radiography using CsI-flat panel detector (FPD) was investigated in consideration of image quality and patient dose. The human chest phantom with inserted imitated nodules (diameter: 10 mm, CT value: +30 Hounsfield unit (HU), -375 HU, -620 HU) was used for the measurement of contrast-to-noise ratio (CNR) of imitated nodules by twenty beams arranged by five tube voltages and four filters. The CNR varies with X-ray tube voltage and added filter. CNR correlates weakly to the tube voltage, fairly to the effective energy in second-order polynomial and strongly to the quality index (effective energy divided X-ray tube voltage). In order to improve the CNR, the effective energy and the quality index are kept about 50 keV and more than 0.5, respectively, using an 80-100 kV beam with a copper filter. A 90 kV (2.5 mm Al inherent filtration) beam with a 0.15 mm copper filter and a 90 kV or 100 kV (2.5 mm Al inherent filtration) beam with a 0.2 mm copper filter are appropriate for chest radiography using CsI-FPD.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have