The combination of interband cascade lasers (ICL) with cavity-enhanced absorption spectroscopy (CEAS) offers new perspectives in trace analysis and isotope ratio measurements. ICLs cover a mid-infrared spectral window (3–4 µm), in between those covered by Ga(InAs)Sb diode lasers and quantum cascade lasers (QCL), where strong molecular transitions can be found. While ICLs have lower emission power than QCLs, their thermal dissipation is much closer to that of telecom diode lasers and their current tuning range larger, which are both major advantages for developing compact instruments. We present an OF-CEAS implementation with an ICL at 4.015 µm, in which optical feedback (OF) enables efficient injection into the high-finesse cavity. In this paper, we also discuss a procedure allowing to obtain an accurate measurement of the OF rate. With regard to performance, we obtain a rms noise-equivalent absorption of 7.7 × 10−9 cm−1 for one acquired spectrum (80 ms) with a cavity of finesse 3900, which translates to a normalized figure of merit of 2.2 × 10−9 cm−1/√Hz, allowing for SO2 trace analysis down to ppbv levels with a response time of seconds.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.