Abstract

Optical coherence tomography angiography (OCTA) can provide in vivo three-dimensional microvasculature information of bio-tissues, but it is sensitive to motion and time-consuming. To overcome these limitations, we propose an adaptive multiple time interval correlation mapping OCTA with a time-efficient scanning protocol and motion compensation algorithms. A spectral-domain OCT with a center wavelength of 850 nm, A-scan rate of 120 kHz and spatial resolution of 4.1 μm (axial) × 6.9μm (lateral) is built to reconstruct the microvascular networks in the human arm. By adaptive optimization of the weights of different time interval B-scan angiograms, our novel OCTA technique achieves better performance with a visible vascular density increase of ~67% and a signal-to-noise ratio enhancement of ~11.6%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.