Abstract
In this work, the possibility of application of free base porphyrin as a lipophilic pH chromoionophore for the preparation of optical cation-selective sensors was investigated. The properties of polymeric membranes, containing porphyrins of different structures, namely tetraphenylporphyrin (TPP) and octaethylporphyrin (OEP), were compared. Changes in equilibrium between protonated and deprotonated form of porphyrin, resulting from variations in ACh concentration, were evaluated. The influence of various factors (kind and quantity of anionic additive and porphyrin in the membrane phase, pH of sample solution) on initial equilibrium was studied. The best membrane composition was chosen as: TPP 3 wt.%, KTFPB 175 mol.% relative to ionophore, PVC:o-NPOE (1 : 4) and measuring buffer solution: 0.05 M MES, pH 4.5. Selectivity, response stability, reversibility and repeatability tests were carried out for chosen sensor. Developed sensor allowed for the determination of a model analyte, acetylcholine, at the concentration range of 10(-5) to 10(-2) M, both in stationary and flow-injection system. Sensor response was reversible and repeatable in the mentioned concentration range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.