Abstract
The eye has a remarkably complex structure and biomechanics. An imbalance between the forces acting on several muscles and the tissue properties may alter or even preclude vision. With the widespread use of electronics, refractive errors have become a common problem, making clear fundus imaging difficult. To overcome this challenge, freeform surfaces have emerged as a potential solution, enabling compact, low-cost, and user-friendly systems that are insensitive to refractive errors. We propose a fundus camera based on a freeform surface that can image the fundus without the influence of refractive errors. A special front lens was designed to image the pupil on a freeform lens that can realize phase modulation. The system performs well providing a field of view of 40∘ and a pupil diameter of 3 mm for refractive errors ranging from -5D to +5D. The volume of the system was less than 35 mm × 35 mm × 135 mm.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.