Abstract
For the first time a segmented n-type HPGe detector was operated directly submerged in liquid nitrogen over a long period. As this kind of detector is envisioned to be used in GERDA phase II, it was operated with a low mass signal cable with snap-contacts and mounted in a low-mass copper holder. The detector performance was stable over 146 days, indicating that neither detector nor contacts deteriorated.