A high-capacity poly(vinyl chloride) flow-through cell which can be used at a flow rate of 3000 ml min -1 in a corrosive environment is described for the continuous on-line monitoring of copper ions in plant electrolytes. The copper(II) ion-selective electrode and double-junction reference electrode in the cell are coupled to a voltmeter and microprocessor-controlled instrumentation to achieve on-line monitoring. The system is designed to withstand the severe industrial environments and was field-tested in a copper refinery where continuous monitoring of copper is required. Cell design and results of the investigation on the determination of copper in plant electrolytes are described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.