AbstractA novel method of measuring the concentration of the local anesthetic ropivacaine drip bags used in hospi‐ tals was developed using electrochemical impedance technique. Polypyrrole (Ppy)/graphene oxide (GO) composites were prepared by electrochemically polymerization over Au electrodes, which served as working electrodes. The Ppy/GO composite electrodes were characterized using scanning electron microscopy (SEM) and transmission electron microscopy (TEM). Various concentrations of ropivacaine (0.1–20 ppm) were prepared in a sample drip bag solution of 0.9 % NaCl from hostpital. The composite Ppy/1 % GO electrode exhibited strongest linearity (R2=0.960) than the Ppy electrode (R2=0.928) at a frequency of 100 kHz. The detection limits of ropivacaine on Ppy and Ppy/1 % GO were calculated as 0.11 ppm and 0.08 ppm, respectively. The proposed system yielded response and recovery times under 1 sec detecting 0.1 ppm (100 g/L) of ropivacaine; thus, the proposed online method is promising for measuring ropivacaine concentrations in drip bags. Molecular simulations and equivalent circuits were applied to explain the dynamic behavior of ropivacaine detection system using Ppy and Ppy/GO materials.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.