Abstract

Office-based endoscopic laser surgery is an increasingly popular option for the treatment of many benign and pre-malignant tumors of the vocal folds. While these procedures have been shown to be generally safe and effective, recent clinical studies have revealed that there are a number of challenging locations inside the larynx where laser light cannot be easily delivered due to line-of-sight limitations. In this paper, we explore whether these challenges can be overcome through the use of side-firing laser fibers. Our study is conducted in simulation, using three-dimensional models of the human larynx generated from X-ray microtomography scans. Using computer graphics techniques (ray-casting), we simulate the application of laser pulses with different types of laser fibers and compare the total anatomical coverage attained by each fiber. We consider four fiber types: a traditional "forward-looking" fiber - not unlike the ones currently used in clinical practice - and three side-firing fibers that emit light at an angle of 45, 70, and 90 degrees, respectively. Results show that side-firing fibers enable a ~70% increase in accessible anatomy compared to forward-looking fibers.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call