Abstract

The Channelized Hotelling observer (CHO) is well correlated with human observer performance in many CT detection/classification tasks but has not been widely adopted in routine CT quality control and performance evaluation, mainly because of the lack of an easily available, efficient, and validated software tool. We developed a highly automated solution - CT image quality evaluation and Protocol Optimization (CTPro), a web-based software platform that includes CHO and other traditional image quality assessment tools such as modulation transfer function and noise power spectrum. This tool can allow easy access to the CHO for both the research and clinical community and enable efficient, accurate image quality evaluation without the need of installing additional software. Its application was demonstrated by comparing the low-contrast detectability on a clinical photon-counting-detector (PCD)-CT with a traditional energy-integrating-detector (EID)-CT, which showed UHR-T3D had 6.2% higher d' than EID-CT with IR (p = 0.047) and 4.1% lower d' without IR (p = 0.122).