Abstract

The paper presents a readout circuit associated with photodiodes dedicated to an in-vivo dosimeter. The circuit consists of a fully differential charge sensitive amplifier (CSA) stage, a differential-to-single-ended amplifier stage and a buffered double sampling stage. The CSA implements a switched integration method to minimize dark current contribution to integrated output signal. Its fully differential operation rejects charge injection effect and further cancels dark current effect. Noise analysis is performed to identify main parameters for low-noise design. The double sampling stage is operated by non correlated double sampling (NCDS) to effectively reduce 1/f noise. The circuit operation is validated by transistor-level simulations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.