Abstract
The basal plane pyrolytic graphite (BPPG) electrode is widely used as a “standard” electrode for studies of electrochemical activity of graphite, graphene and carbon nanotubes. We compared here different methods for preparation of the surface of BPPG using statistical analysis of the data. We showed that including/omitting of polishing step or including/omitting of acetone washing step in the preparation procedure exhibits statistically significant difference. However, using acetone of different purities does not lead into statistically significant results. In addition, all methods with BPPG provided RSD < 10% and did not produce any outlying results.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.