Abstract

A special self-designed device based on poly-1-vinyl-3-pentylimidazole hexafluorophosphate (PILs-C5) solid-phase extraction and high performance liquid chromatography (HPLC) is proposed as a novel method for the on-line separation and analysis of Rhodamine B (RhB) dye. Single factor experiment design and orthogonal experiment design were used to optimize the experimental parameters, such as pH, the amount of PILs-C5, sample volume, flow rate, eluent type, eluent concentration, eluent volume, and the flow rate of eluent. Under the optimized conditions, the linear range was 0.02–2.4 μg mL−1, with the correlation coefficients (R2) of 0.997. The limit of detection (LOD) and limit of quantification (LOQ) were 0.004 μg mL−1 and 0.02 μg mL−1, respectively. The extraction capacity was 6.22 mg g−1, and enrichment ratio was 15. The extraction mechanism and the post-treatment method of PILs-C5 were also studied. This method was applied to analyze RhB in a wide variety of real samples with satisfactory results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.