Abstract

A flow-injection biosensor system with an on-line microdialysis sampling is proposed for the simultaneous assay of l-lactate and pyruvate in serum and rat brain. The dialysate collected in the sample loop by perfusing Ringer’s solution through the microdialysis probe is automatically injected into the flow-injection line with a dual enzyme electrode arranged in parallel for the flow direction. The dual enzyme electrode is constructed by hybridizing a poly(1,2-diaminobenzene) film to two sensing parts, which respond selectively to l-lactate and pyruvate, respectively, without any cross-reactivity. Both the sensing parts respond linearly to the concentrations of both analytes between 0.01 and 5mM, without any interference from oxidizable species and low-molecular weight proteins present in the dialysate. The proposed flow-injection analysis (FIA) method can be successfully applied to the simultaneous in vitro and in vivo assays of both analytes in serum and rat brain, respectively. The system can be automatically processed at an analytical speed of 19dialysatesh−1 over a period of 5h.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call