Abstract

A reversed-phase high-performance liquid chromatography-mass spectrometry (LC-MS) method is described for the separation and simultaneous analysis of porphyrins related to disorders of heme biosynthesis (uro-, heptacarboxylic, hexacarboxylic, pentacarboxylic, and coproporphyrins). The method involves initial porphyrin esterification and extraction from urine. Detection and quantification is performed from the extracts by separation with a Hypersil BDS column and on-line detection by MS through coupling with an atmospheric pressure chemical ionization interface. The porphyrin esters are detected as protonated molecules [M + H]+. Their mass spectra also exhibit an [M + Na]+ fragment of lower intensity. The analytical performance of this method is compared with those of LC with UV and fluorescence detection. LC-MS used in selective [M + H]+ ion monitoring provides the lowest detection and quantitation limits. In scan mode, this LC-MS method affords, without further isolation or concentration steps, the measurement of mass spectra of unknown compounds present in the urine of patients with altered porphyrin excretion.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.