Abstract

A new method has been developed for increasing the sensitivity of collinear laser spectroscopy. The method utilizes an ion-trapping technique in which a continuous low-energy ion beam is cooled and accumulated in a linear Paul trap and subsequently released as a short (10-20 micros) bunch. In collinear laser measurements the signal-to-noise ratio has been improved by a factor of 2 x 10(4), allowing spectroscopic measurements to be made with ion-beam fluxes of approximately 50 ions s(-1). The bunching method has been demonstrated in an on-line isotope shift and hyperfine structure measurement on radioactive (175)Hf.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.