Abstract
A fully automated HPLC method for analysing the non-ionic X-ray contrast agent iodixanol in plasma samples, using on-line dialysis for sample preparation, was developed. Optimal conditions were obtained with a static dialysis donor solution of 110 μl and 4 ml of recipient solution (dialysate) pulsed onto a trace enrichment column, giving maximum 55% dialysis efficiency in less than the chromatographic run time of 20 min. Hence, one sample could be dialysed during the analysis of the previous. The maximum load of iodixanol on the trace enrichment column was 3.75 mg. Validation showed that the method was selective for iodixanol, sensitive down to 84 pmol/ml and had a high precision over a linear range up to 320 nmol/ml. Although developed for iodixanol, the method can easily be modified and applied to other substances with similar properties, i.e., substances having low protein binding and high water solubility, but strong enough stationary phase affinity to be retained by an appropriate trace enrichment column.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.