Abstract

A simple and rapid procedure for SO2 determination in air was developed by using a flow injection analysis (FIA) system coupled with a 3-hole chromatomembrane cell (CMC). The CMC was applied for the on-line collection/concentration of SO2 from air into a solution of 2 g l(-1) triethanolamine (TEA) solution as an absorbing solution: SO2 was converted to SO3(2-) in the alkaline absorbing solution. The solution containing absorbed SO2 was introduced into the carrier stream of the FIA system. The amount of SO3(2-) in the absorbing solution was measured by spectrophotometry with a mixed reagent of pararosaniline and formaldehyde, and was converted to the concentration of SO2 in the air sample. A calibration graph prepared by using standard sodium sulfite aqueous solutions was adopted for the determination of SO3(2-) in the absorbing solution. The SO2 concentration in indoor air examined was found to be 22.7 +/- 0.2 ppbv using 20 ml of air sample with the air flow rate of 5 ml min(-1), where the relative standard deviation was 1.7%. The detection limit for aqueous solutions and air samples were 6.9 x 10(-8) M and 0.48 ppbv, respectively. The measuring time for one sample was about 10 min when a 20 ml air sample was used. The interferences from common anionic species, formaldehyde and acetaldehyde, were also examined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.