Abstract
Abstract Thermally-assisted laser induced fluorescence temperature measurements using upward vibrational transfer from υ' = 0 to υ' = I in A2 Σ+ OH have been demonstrated in premixed nitrogen diluted flames of acetylene, ethylene, and methane over a wide range of equivalence ratios. Detection of the (0-0) and (1-0) vibrational bands was used to determine the relative populations of υ' =0 and I from which temperature was calculated assuming a simple Iwo level model incorporating the effects of vibrational energy transfer and quenching. It was found that collisional effects that are a source of systematic error can be accounted for through a calibration procedure for each fuel type. The resulting accuracy of the measurement is better than I00K as determined through comparison with sodium line reversal temperature measurements in flames that ranged in temperature from 1700 K to 2700 K. Similar results using detection of the (0-1) and (1-0) vibrational bands in order to minimize self-absorption effects were a...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.