Abstract
Octadecylsilyl-bonded silica based novel solid phase extraction membranes were fabricated by non-solvent vapor induced phase separation method for the first time, and employed for extraction trace phenol as a model. The effects of membrane fabrication conditions and extraction parameters on the recoveries were explored. Under the optimized fabrication conditions, the extraction recovery of phenol (0.02 mg/L, 100 mL) attains 100% with enrichment factor of 10. The spike recovery of real water samples reaches 97.67%–99.40% with standard deviation of 1.16%–1.42%, demonstrating great potential in solid phase extraction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.