Abstract

To assess exposure of MR staff to static and time-varying gradient magnetic fields by direct measurement and questionnaire and to compare these to the action values and exposure limits stated in published guidance. Instantaneous and time-averaged exposure to static magnetic fields were measured on clinical and research MR scanners. A questionnaire was sent to 100 clinical MR centers to estimate the average static field exposure. Instantaneous measurements of the time-varying magnetic field were conducted on a 1.5 T conventional and a 0.6 T open-bore MR system. The 24-hour time-averaged static field exposure was measured to be 1.8 +/- 1 mT and the questionnaire produced a 24-hour time-average of 2.3 mT. The maximum static field strength recorded during patient set up was 554 mT. Time-varying gradient field exposure was found to exceed the action values up to 30 cm from the bore entrance or magnet rim. The time-varying gradient field exposure limit is the only limit to be exceeded. The time-averaged static field exposure to staff was found to be 100 times lower than the recommended guidelines.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.