Abstract

Arterial tonometry is a widely used non-invasive blood pressure measurement method. In contrast to the cuff-based method, it is possible to obtain a continuous pressure profile with respect to systolic and diastolic pressures using this method. However, due to a requirement of arterial tonometry—that a sensor needs to be placed directly above a blood vessel—placement error is inevitable if the measurement device is only capable of measuring local regions. This study assumed that the plate sensor is flexible, thus reducing the placement error. We investigated the pressure distribution along the wrist surface rather than the local region through the contact simulation between the flexible plate sensor and the wrist. As a result, we concluded that there is a unique pressure distribution for any specific wrist, regardless of the length and position of the plate, and that it is possible to measure the blood pressure using the response at the wrist surface to the pressure inside the radial artery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.