Abstract

To identify a novel material with biomechanical properties identical to the nucleus pulposus in the lumbar vertebrae of goats for use in intervertebral disc herniation surgery. In this laboratory-based experimental study, the silicone rubber material test group consisted of sample compositions 35PVA65SR, 30PVA70SR, and 40PVA60SR. Axial compression mechanical tests were conducted to assess the biomechanical properties of the resulting material in terms of stress, strain, load, and displacement. The mechanical compression test results revealed that the stress (MPa) and strain (%) values of the 40PVA60SR material were closest to the control group (p=1.00) with a load of 684 N in each group. The value of material displacement (mm) for 40PVA60SR was also the closest to control (p=1.00) with a loading of 684 N. The new material presents biomechanical properties closest to human nucleus pulposus and is promising in nucleus pulposus replacement therapy. Further clinical research is needed to evaluate other biomechanical properties and the bioavailability of the novel material.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.