Abstract

A mixture of manganese dioxide and hydrochloric acid has been used for the accurate and precise determination of gold in various gold-bearing matrices. Results of intermethod comparison studies, F-test on variances, Mann-Whitney U-test, Spearman rank correlation and regression analyses are presented. The recommended method can be applied to a quality control programme and for the evaluation of reference materials. Various standard reference materials (SRM) of diverse matrices have been examined to check the validity of the method and the results were found to be in very good agreement with the certified data. Sample decomposition is straight forward [1–10 g sample (roasted at 600–700 °C) + 5 g of MnO2 + 30 mL of 2 mol/L HCl]. Gold is coprecipitated with potassium tellurite using tin (II) chloride solution. The precipitate is extracted into toluene and finally stripped back into aqua-regia solution for final nebulization into an airacetylene flame for atomic absorption spectrophotometry. The method is very simple and easily adaptable, and more convenient than conventional methods involving aqua regia or a hydrobromic acidbromine water mixture.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.