Abstract

We have developed a planar chip utilizing divergent geometry of separation channel capable of vertical free-flow electrophoresis of particles at flows of lower hundreds of microliters per minute. The divergent flow isoelectric focusing (DF-IEF) chip consists of two sheets of clear polystyrene glass which serve as a base with working channels and a top cover sealing the separation channel. Optimization showed that the chip is capable to form pH gradient within 1 h and separation is completed in 5 or more minutes depending on the sample volume. The vertical position of the chip enabled analysis of sedimenting particles including microorganisms. Four different common bacteria species inactivated with H2O2 vapors were analyzed in a series of experiments. Isoelectric points were determined with capillary isoelectric focusing with following fractionation using DF-IEF with intact cell matrix-assisted laser desorption/ionization mass spectrometry detection. The DF-IEF chip fractionation proved promising for bacterial sample preparation from complex matrices for subsequent identification of whole cells by mass spectrometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.