Abstract

Switched capacitor arrays (SCA) ASICs are becoming more and more popular for the readout of detector signals, since the sampling frequency of typically several gigasamples per second allows excellent pile-up rejection and time measurements. They suffer however from the fact that their sampling bins are not equidistant in time, given by limitations of the chip process. In the past, this limited time measurements of optimal signals to standard deviations ({\sigma}) of about 4-25 ps in accuracy for the split pulse test, depending on the specific chip. This paper introduces a novel time calibration, which determines the true sampling speed of an SCA. Additionally, for two independently running SCA chips, the achieved time resolution improved to less than 3 ps ({\sigma}) independently from the delay for the split pulse test, when simply applying a linear interpolation. When using a more advanced analyzing technique for the split pulse test with a single SCA, this limit is pushed below 1 ps ({\sigma}) for delays up to 8 ns. Various test measurements with different boards based on the DRS4 ASIC indicate that the new calibration is stable over time but not over larger temperature variations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.