Abstract
The focusing effects due to fringing fields at both the entrance and the exit slits of the cylindrical mirror energy analyser are examined and a set of corrected design parameters is presented which takes account of the second order focusing in the cylindrical field. The resolution is rather poorer in this case compared with the 'classical' case; however transmission losses as well as secondary emission at the window grids are avoided.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.