Abstract

BackgroundQuantitative T1-mapping is rapidly becoming a clinical tool in cardiovascular magnetic resonance (CMR) to objectively distinguish normal from diseased myocardium. The usefulness of any quantitative technique to identify disease lies in its ability to detect significant differences from an established range of normal values. We aimed to assess the variability of myocardial T1 relaxation times in the normal human population estimated with recently proposed Shortened Modified Look-Locker Inversion recovery (ShMOLLI) T1 mapping technique.MethodsA large cohort of healthy volunteers (n = 342, 50% females, age 11–69 years) from 3 clinical centres across two countries underwent CMR at 1.5T. Each examination provided a single average myocardial ShMOLLI T1 estimate using manually drawn myocardial contours on typically 3 short axis slices (average 3.4 ± 1.4), taking care not to include any blood pool in the myocardial contours. We established the normal reference range of myocardial and blood T1 values, and assessed the effect of potential confounding factors, including artefacts, partial volume, repeated measurements, age, gender, body size, hematocrit and heart rate.ResultsNative myocardial ShMOLLI T1 was 962 ± 25 ms. We identify the partial volume as primary source of potential error in the analysis of respective T1 maps and use 1 pixel erosion to represent “midwall myocardial” T1, resulting in a 0.9% decrease to 953 ± 23 ms. Midwall myocardial ShMOLLI T1 was reproducible with an intra-individual, intra- and inter-scanner variability of ≤2%. The principle biological parameter influencing myocardial ShMOLLI T1 was the female gender, with female T1 longer by 24 ms up to the age of 45 years, after which there was no significant difference from males. After correction for age and gender dependencies, heart rate was the only other physiologic factor with a small effect on myocardial ShMOLLI T1 (6ms/10bpm). Left and right ventricular blood ShMOLLI T1 correlated strongly with each other and also with myocardial T1 with the slope of 0.1 that is justifiable by the resting partition of blood volume in myocardial tissue. Overall, the effect of all variables on myocardial ShMOLLI T1 was within 2% of relative changes from the average.ConclusionNative T1-mapping using ShMOLLI generates reproducible and consistent results in normal individuals within 2% of relative changes from the average, well below the effects of most acute forms of myocardial disease. The main potential confounder is the partial volume effect arising from over-inclusion of neighbouring tissue at the manual stages of image analysis. In the study of cardiac conditions such as diffuse fibrosis or small focal changes, the use of “myocardial midwall” T1, age and gender matching, and compensation for heart rate differences may all help to improve the method sensitivity in detecting subtle changes. As the accuracy of current T1 measurement methods remains to be established, this study does not claim to report an accurate measure of T1, but that ShMOLLI is a stable and reproducible method for T1-mapping.

Highlights

  • Quantitative T1-mapping is rapidly becoming a clinical tool in cardiovascular magnetic resonance (CMR) to objectively distinguish normal from diseased myocardium

  • Cardiac T1-mapping without the use of exogenous contrast agents has been shown to be sensitive to a variety of pathologies, notably acute myocardial infarction [3,4], myocarditis [5] and cardiac amyloidosis [6], all of which demonstrate significant (i.e. 20-30%) increase in native T1 times

  • We examined the effects of normal physiologic parameters, including age, gender, heart rate, weight, height and hematocrit, as well as technical factors, such as partial volume effects, myocardial thickness and inter-centre variability on identical scanners within and across three clinical centres in two countries

Read more

Summary

Introduction

Quantitative T1-mapping is rapidly becoming a clinical tool in cardiovascular magnetic resonance (CMR) to objectively distinguish normal from diseased myocardium. The usefulness of any quantitative technique to identify disease lies in its ability to detect significant differences from an established range of normal values. Each tissue type, including myocardium, exhibits a characteristic range of normal T1 relaxation times at a particular magnetic field strength [2], and deviation from the normal range may be indicative of disease. In order to use native myocardial T1-mapping to accurately identify disease states, a stable normal range and factors that influence it need to be established. These include common physiologic variation in the normal population as well as technical factors inherent to the measurement technique and image analysis

Objectives
Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.