Abstract
In some diagnoses, such as the polysomnography, simultaneous measurement of the electroencephalogram (EEG) and the electrocardiogram (ECG) is often required. It would be more efficient if both the EEG and ECG could be obtained simultaneously by using a single measurement. In this paper, we introduce a nonlinear state-space projection-based technique to extract the EEG and ECG components from an EEG signal measured with a non-cephalic reference (NCR) that guarantees accurate detection of R waves in the EEG measurement. Evaluation of the method using simulated data showed that the improved normalized power spectrum in alpha, beta (13-30 Hz), and theta bands were accurate. In an accrual EEG, measured using the NCR electrode, it was confirmed that the frequency components of the extracted EEG were accurate, and no spikes that could be attributed to the ECG component were observed in the resultant EEG signal.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Life Science and Medical Research
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.