Abstract

Blood pressure was predicted from photoplethysmography (PPG). To obtain PPG, backscattered light from a fingertip was measured and its waveform was analyzed. Systolic upstroke time and diastolic time in the pulse waveform were used as parameters to predict blood pressure. The experiment was carried out with five subjects on five different days. The systolic upstroke time had a correlation coefficient of -0.605 with respect to systolic blood pressure and the diastolic time had a correlation coefficients of -0.764 for diastolic pressure. This PPG method does not require an air-cuff installation on the arm and can predict blood pressure continuously. This simple LED/photo detector setup can be a good candidate for nonconstrained monitoring of blood pressure variations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.