Abstract
This paper investigates a method for conducting conventional ultrasound image scanning (C-scan) in a non-immersive manner. An ultrasound coupling device is developed to couple a focusing ultrasound transducer with a sample under test through a water droplet. The transducer is immersed in a chamber filled with water. A hole-opening at the chamber’s bottom surface and a polymer film sealing the hole allow focused ultrasound to pass and reach the sample through a water droplet. An ultrasound scanning system is constructed and used for carrying out C-scan on two different samples containing pre-fabricated internal defects. By comparing the images obtained from using both conventional water immersion method and the proposed water-droplet coupling method, we demonstrate the feasibility and capability of this new type of non-immersive ultrasound image scanning method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.