Abstract

We aimed to evaluate the image quality of brain computed tomography (CT) images reconstructed using deep learning-based reconstruction (DLR) in organ-based tube current modulation (OB-TCM) acquisition. An anthropomorphic head phantom and a cylindrical low-contrast phantom were scanned at the standard dose level for adult brain CT in axial volume acquisition without OB-TCM. Moreover, image acquisition with OB-TCM was performed. The radiation dose on the eye lens was measured using a scintillation fibre-optic dosimeter placed on the anthropomorphic phantom's eye surface. The task transfer function (TTF), contrast-to-noise ratio (CNR), and low-contrast object specific CNR obtained from low-contrast phantom images reconstructed with filtered back projection (FBP), hybrid iterative reconstruction (HIR), and two types of DLR (DLRCTA and DLRLCD) were compared. In result, OB-TCM achieved a 32.5% dose reduction in the eye lens. Although HIR, DLRCTA, and DLRLCD showed lower TTF than FBP, the difference in TTF at the highest contributing spatial frequency corresponding to the contrast rod diameter was < 10%. Despite the OB-TCM acquisition, DLRCTA and DLRLCD achieved significantly lower noise and a higher CNR than FBP without OB-TCM (p < 0.05). However, low-contrast object specific CNR was equivalent among all reconstruction methods for the objective diameter of 5mm and slightly improved in DLRLCD for the objective diameter of 7mm. DLR with OB-TCM acquisition enabled dose reduction for the eye lens and high CNR image appearance, whereas the low contrast detectability evaluated by low-contrast object specific CNR did not always improve.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call