Abstract

The purpose of this study was to assess tibial post fracture between highly cross-linked polyethylene (HXPE) and conventional polyethylene inserts in consecutive posterior-stabilized (PS) total knee arthroplasty (TKA) over a minimum 10-years follow-up period. Between July 2008 and December 2011, a consecutive series of 3264 primary TKAs was performed in 2098 patients using a PS total knee prosthesis at a single institution. The final cohort was classified into the HXPE group (792 cases) and the conventional group (2113 cases) according to insert material used during the procedure. The mean follow-up period was 11.4years in the HXPE group and 11.8years in the conventional group. Both cases of tibial post fracture occurred in the HXPE group, a complication rate of 0.25%. These two cases required replacement with a thicker tibial insert with retention of the femoral and tibial components. There were no cases of tibial post fracture in the conventional group. There was no difference in tibial post fracture rate between the HXPE group and the conventional group in PS TKA on long-term follow-up. IV.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call