Abstract

On use of nitrous oxide as carrier gas the retention factors of the chromatographed compounds decrease linearly with increasing average column pressure. Other retention characteristics (relative retention, retention index) change linearly. This effect was demonstrated by using a capillary column coated with nonpolar polydimethylsiloxane phase SE-30. As shown for capillary GLC the linear correlation is valid for the same column:ki(G1,P1) = A ki(G2, P2) + B, where ki(G1, P1) and ki(G2,P2) are the retention factors of compound i at average column pressures P1 and P2 when using carrier gases G1 and G2, respectively; A and B are coefficients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.