Abstract

The need for on-line sample preparation for high-throughput applications in bioanalysis has increased during the past decade. In this paper a robust and on-line sample preparation technique, micro extraction in packed syringe (MEPS) has been developed and validated. The method is a miniaturized, fully automated, solid-phase extraction (SPE) technique that can be connected on-line to GC or LC without any modification of the chromatographs. The performance of MEPS as sample preparation method is illustrated by the determination of local anaesthetics in human plasma samples on-line with high performance liquid chromatography (HPLC) and tandem mass spectrometry. The sampling sorbent was 1 mg silica based benzenesulphonic acid cation exchanger that was inserted in a 250 μl syringe. Ropicavine and two of its metabolites (PPX and 3-OH-ropivacine), lidocaine and bupivacine were used as model substances. The accuracy values of quality control samples (QC) were between 95% and 109%, and precision (relative standard deviation, R.S.D.) had a maximum deviation of 9% for the analytes.