Abstract

To improve the performance of mono-extruded TLD threads as a dosimetric thermoluminescent tool (French Patent 9903729), a new process was developed by co-extrusion methodology leading to threads of 600 µm diameter with a 50 µm homogeneous polypropylene sheath. In this optimization work, study of parameters such as LiF:Mg,Cu,P powder granulometry, load rate and proportion of components led to an increased sensitivity of around 40%. Moreover, the co-extrusion technique allowed the threads to be sterilized by humid steam (134 °C/18 min) without significant variation of the linearity response between 0 and 30 Gy after gamma irradiation (60Co).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.